Power frequency EMF measurements

John Swanson, Technology. & Science Labs., National Grid Transco Research. & Development. Centre, Leatherhead, UK

1              General principles

Measurements of electric and magnetic fields can be made for various purposes.  Depending on the purpose, different techniques will be appropriate.  There is no single “correct” procedure; the correct procedure will depend on the reason the measurements are being made and the use that will be made of the results.

 

When embarking on a measurement programme, therefore, it is important first to identify clearly the objective.  This objective must be used to decide on the procedures to be followed.  This chapter discusses the choices and issues involved, and gives suggestions for appropriate procedures, but is not a substitute for intelligent, informed choice on the part of the person specifying or performing the measurement.

 

The main international standard relating to measurements at power frequencies is a 1998 IEC standard IEC-61786, “Measurement of low-frequency magnetic and electric fields with regard to exposure of human beings – Special requirements for instruments and guidance for measurements”.  The provisions of this standard are referred to where appropriate in the following material.

2              Types of measurement

The following are some of the main reasons for performing measurements.  This is not, however, an exhaustive list.

 

Simple characterisation of the field in a building

The objective is to characterise the field in a home, a work location, or similar, by a single number.  Depending on resources and time available, this can be done by a single measurement at a single location; a single measurement at each of a series of locations; a sequence of measurements over time at a single location; or a sequence of measurements over time at multiple locations.

 

Identification of sources

The objective is to make measurements of the field at a location, specifically how it varies over space or time, to enable the source of the field to be identified.

 

Comprehensive characterisation of the field in a building

The objective is to collect more data on the field in a building than just a single number, so as to permit extraction of desired information at a later date.

 

Characterisation of sources

The objective is to perform measurements that relate to a particular source of field (eg a power line, an item of equipment) rather than to the field in a particular place, so as to characterise that source.

 

Personal exposure

The objective is to measure the exposure of a person over a period of time during which they are exposed to fields from various sources.

 

Compliance with exposure limits

The objective is to assess whether a given set of EMF exposure limits are exceeded.  This can be done either by assessing the fields in an area in which people will be present, or by monitoring the exposure of the people, or a combination.

 

Laboratory measurements

The objective is to measure the field within experimental apparatus in a laboratory, eg the field produced by coils used to expose an experimental system to magnetic fields.

 

3              Choice of instrument for magnetic field measurements

 

3.1        Sensor technology

There are three technologies for measuring magnetic fields: fluxgate magnetometers, Hall effect devices, and search coils.

3.1.1.                                                      Fluxgate magnetometers

These comprise a ferromagnetic core which is driven in and out of magnetic saturation in opposite directions by a high-frequency current generated within the instrument.  The addition of an external field creates an asymmetry, and the instrument uses this asymmetry to measure the field.  Fluxgate magnetometers are sensitive to all external fields up to a certain frequency, static fields as well as alternating fields.  The static and alternating components of the field can be separated and recorded separately by the instrument circuits.

 

Fluxgate magnetometers are usually the preferred choice when it is desired to measure static fields as well as alternating fields.  The fluxgate sensor itself can be made reasonable small and is usually remote from the bulkier remainder of the instrument, so they can have uses where it is necessary to probe the field in difficult locations.  Other than these specific applications, their use is limited, as they are more expensive and have higher battery consumption than search coil instruments.  Care should be taken if using them in laboratory settings, as they produce a finite high-frequency field themselves as part of their operation which could perturb an experimental setup.

 

3.1.2.                                                     Hall effect devices

These devices pass a current through a suitable semiconductor, and detect the field via the voltage produced across the element perpendicular to the current.  Like fluxgate magnetometers, they detect the instantaneous total field, the sum of the static and any alternating fields.  They are often used in other applications because they can measure high fields at the Tesla level and above.  However, they are usually not very sensitive to low fields and suffer from zero-point drift, and the probes are often fragile, which means they have few applications in EMF dosimetry.

 

3.1.3.                                                     Search coils

Search coils are simply coils of wire.  An alternating magnetic field induces a voltage in the coil.  Except for the specialised instances where fluxgate magnetometers are appropriate, search coils are the preferred technology for EMF dosimetry where it is not desired to measure the static field as well.

 

Characteristics of search-coil instruments

 

·               Size of coil

To obtain a large enough signal, search-coil instruments tend either to have relatively large coils (10-20 cm), or small coils (less than 1 cm) with ferrous cores.  Larger coils are perfectly acceptable as long as the field itself does not vary significantly over the area of the coil.  Thus they will often be acceptable under power lines or in the middle of rooms, but will be less appropriate close to conductors or to equipment.  IEC specifies that the coil should have area 0.01 m2 or less.

 

It is sometimes recommended to use large coils so as to measure the average field over an area comparable to part of the human body.  Usually, however, where the field varies over space this rapidly, it is preferable to measure the variation of the field with a small probe and then to apply any desired averaging to the results of those measurements.

 

Small, ferrous-cored coils are often preferred and produce a smaller, more versatile instrument.  The one disadvantage is that the ferrous core produces non-linearities at high fields, greater than 1 mT.  For most purposes this is irrelevant as fields are rarely that high, but care should be exercised if high fields are to be measured.

 

·               Number of coils

Each coil measures the component of field in one direction.  Instruments have either one or three orthogonal coils.  The choice depends on the purpose of the measurement.  When identifying and investigating different sources of field, a single coil can be useful, as the extra information on the direction of the field can assist in identifying the source.  For most other applications, however, three coils are more useful.

 

The resultant field from multiple sources at the same frequency always traces out an ellipse.  There are two alternative ways of quantifying an elliptically polarised field, shown in figure 1. 

 


Figure 1         Alternative measures of an elliptically polarised field. 1:

maximum” field.  2: “resultant” field

 

Vector 1 gives the rms of the field along the major axis of the ellipse, which is the direction of maximum field.  This is known as the “maximum” of the field.  Vector 2 gives the actual rms of the field, known as the “resultant field”.

 

The “maximum” field, vector 1, can be measured by rotating a single coil until the maximum value is obtained.  The “resultant” field, vector 2, can be measured using three orthogonal coils.  The rms of the signals from the three coils are combined as root-sum-of-squares to give the resultant field.  This applies regardless of the orientation of the coils relative to the field. 

 

Three orthogonal air-cored coils can be arranged so that, to a good approximation, their centres are coincident and therefore they measure the field at the same point in space. Figure 2a shows an example.  Ferrous-cored coils are nearly always physically separate.  This still gives satisfactory results as long as the field does not vary over the distance scale of the separation of the coils.  As this can be small (of the order of a cm) this is usually acceptable.  Figure 2b shows three such ferrous-cored coils.

 



Figure 2  Examples of instruments using

 

(a) three orthogonal coils arranged to be centred on a single point

 

(b) three separate orthogonal coils

If only a single-axis instrument is available, the resultant (total) field can be measured by performing three successive orthogonal measurements.  This can have limited accuracy, however, if the field varies over the time taken to perform these measurements.  It can also be difficult to locate the sensor at the same point each time; if accuracy is required, a jig to locate the instrument should be used.

3.2        Type of display

Increasingly, most displays are digital.  The main occasion when an analogue display is more useful is when investigating a source, when variations in the field as the meter is moved need to be readily observed, but even here a digital display can be used.

3.3        Range and resolution

Average background fields in homes (ie fields in the general volume of the home, not close to equipment) range from a few tens of nanotesla in some European countries to over a hundred nanotesla in North America.  If performing measurements in a high-field country, and particularly if the emphasis of the measurements is on high fields in that country (eg on identifying homes with fields greater than 200 or 400 nT), a resolution of 10 nT will be adequate.  In many other instances for measurements in homes, however, a resolution of 1 nT is desirable.

 

The desired range of the instrument depends on the maximum field likely to be encountered.  Background fields in homes are rarely greater than 1 µT; fields in industrial settings or near domestic equipment can be 1 mT or more.  Occupational exposure limits are typically of the order of a mT, so a range of at least this is necessary to assess compliance with such limits.

3.4        Storage abilities

If using measurements of field to investigate a source, no storage ability is necessary; the changes in field as the instrument is moved are observed in real time.  For some other purposes, storage within the instrument is not needed; the reading can be written down or entered into a database.  For many purposes, however, it is helpful or necessary to store results within the instrument.  The choice of measurement interval and total duration is a compromise limited by the total storage capacity available in the instrument.

3.5        Frequency response

The output of a search coil is proportional to frequency.  Most instruments adapt this frequency response within the instrument to give a final output which is either broadly flat between a lower cut-off (typically 20-30 Hz) and an upper cut-off (typically 500 Hz – 5 kHz), or is sensitive only to the power frequency, 50 or 60 Hz.

 

The choice of frequency response is determined by the characteristics of the field to be measured and the purpose of the measurements.  In many cases, the source of the field is the power system.  Then, the main component of field will be at 50 or 60 Hz, with smaller harmonics at multiples of this, principally three times or “third harmonic”, 150 or 180 Hz.  Most flat responses will be adequate for this.  However, if there is particular interest in harmonics, or if the sources is such that higher harmonics (eg from some ac-dc power conversion processes) or lower frequencies (eg from some railways at one-third the power frequency) are present, then an instrument with a particular frequency response may be appropriate.

 

Note that search coils produce a signal when rotated in a static field.  To avoid this being erroneously recorded as a power-frequency signal, many instruments have filters to reduce the response at low frequencies.  If an instrument sensitive to lower frequencies is used (eg because railways are to be measured), extra care must be taken to avoid this interference.

 

Exactly how flat a frequency response is depends on the sophistication of the instrument.  However, achieving a flatter frequency response also tends to increase the power consumption and the weight of the instrument.  Therefore, instruments designed to be small, light and with long battery life may have poorer frequency responses.  This is part of the choice that has to be made, but in many instances, the flatness of the frequency response will not affect the result much.

3.6        rms and other measures

Increasingly, instruments are made with true rms detection of the field, and this is regarded as preferable for most purposes.

 

Alternatives are rectified average and peak detection.  There may be occasions when each of these is desirable.  However, in general, these are a legacy of less developed instrument technologies, and should be avoided.  They cause particular problems when harmonics are present.

3.7        Accuracy

The accuracy required of an instrument should, in principle, be determined by the measurement purpose.  However, IEC specifies an overall uncertainty in the measurement of a uniform field of ±(10% of reading +20 nT), and most commercial instruments can be regarded as accurate to at least this 10% value.

3.8        Other factors

IEC specifies that an instrument should function over a temperature range of 0 to 45 °C and from 5% to 95% relative humidity.

 

If measurements are to be done in the high electric fields close to high-voltage equipment, the instrument should be immune to these fields.  IEC specifies that fields of 20 kV m-1 should produce no more than 20 nT change in the reading.  Most instruments will have satisfactory EMC performance; IEC specifies immunity to 3 V m-1 from 150 kHz to 1 GHz assessed in accordance with standard IEC 61000 (IEC 1995).

4              Choice of instruments for electric-field measurements

Most practical electric-field measurements are made by measuring the voltage (or current) between two parallel plates perpendicular to the electric field.  Other technologies are available, but have few advantages.  Such a meter is a “free body” meter; the alternative is a “ground reference” meter, which measures the voltage between a single electrode and ground, but this is less common.

 

Parallel-plate sensors for electric fields tend to be physically larger than magnetic-field sensors.  Therefore, electric-field instruments are more often single-axis.  However, three-axis meters are available, with the three separate parallel-plate assemblies either on three orthogonal faces of a cube or spread over the surface of a sphere.

 

One common measurement scenario is near or under power lines.  In this instance, the electric field is essentially vertical for the first metre or two at least above ground level.  Therefore, in this instance, a single-axis meter, aligned to measure the vertical field, is perfectly satisfactory.

 

Many of the issues discussed under magnetic fields apply to measurements of electric fields as well: frequency response, storage, type of display, etc.  There is, however, one major issue with electric fields that does not apply to magnetic fields.

 

4.1        Perturbed and unperturbed electric fields

Electric fields are perturbed by any conducting object.  This includes the person making the measurement and any conducting supports for the sensor.

 

Electric-field measurements should be either of the unperturbed field or of the perturbed field and it should be clear which is being measured.  Measurements made by a meter worn by a person are unavoidably of the perturbed field.  In most other cases, however, it is preferable to attempt to measure the unperturbed field.  This can sometimes be an ambiguous definition, where the field is perturbed by an object not connected with the measurement process.  The usual objective is to measure the field that would be present in the absence of the person, but including any perturbation caused by fixed objects that would still be present when the person is not.

 

To obtain an unperturbed measurement, the instrument sensors must be supported in a way that does not perturb the field, and the person performing the measurement must be sufficiently distant.

 

The instrument can be supported either on an insulating pole held horizontally, or on a vertical insulating support (a tripod or similar).  In both cases, if readings in real time (as opposed to stored and read only later) are desired, the sensors are connected to the rest of the instrument by a radio link or a fibre-optic, or the instrument is self-contained with a display large enough to be read from a distance.

 

A vertical tripod or similar support is often mechanically easier and makes it easier for the operator to be distant.  However, in humid conditions, moisture can settle on both a tripod or a pole, making them less good insulators and perturbing the reading.  As most electric fields are vertical near ground level, the vertical support, which is parallel to the field lines, leads to greater perturbation than the horizontal pole, which runs perpendicular to the field lines, along an equipotential, and therefore is less sensitive to small amounts of moisture.  This is a reason for preferring horizontal poles when some degree of perturbation is regarded as inevitable.

 

The person making the measurements should be 2 m distant from the instrument to ensure that perturbation is negligible.  Measurements at 1 m may be acceptable but less accurate.  Hand-held measurements should be avoided.

 

The perturbation caused by the person making the measurement is, simplistically, to increase the field towards the top of their body, and to reduce it towards the bottom of their body.  There is therefore a height of measurement, around 1.5 m above ground and therefore higher than the 1 m above ground which is recommended as a suitable measurement height for other reasons, where the perturbing effect of the person is neutral.  This means in practice, the distance of the person from the instrument may not be so critical.  However, this should not be relied on, and for good measurements, the person should always be 2 m distant.

 

Measurements of the perturbed electric field are made by attaching a meter to a person’s body.  However, the reading obtained is extremely dependent on the amount of perturbation, which in turn is extremely dependent on the exact location of the meter on the person’s body.  For certain simple geometries and for certain locations on the body (eg the top of the head of a person standing upright in a vertical field, where the factor is approximately 20) it is possible to calculate the conversion factor from the perturbed to the unperturbed field.  In most situations, however, this conversion factor cannot be calculated, and varies as the person moves anyway.  Therefore, such perturbed measurements have limited use, and should never be compared directly to unperturbed fields.

 

The need to measure unperturbed fields undoubtedly adds considerably to the effort and cost of making electric-field measurements, but in most situations, it is the unperturbed rather than the perturbed field which is the most useful.

5              Calibrations

It is necessary to be sure that any instrument used is giving an acceptably accurate reading; this is achieved by calibration.  IEC requires the overall uncertainty of the calibration process to be no greater than ±(5% +10 nT).

 

Most instruments will be supplied calibrated by the manufacturer.  Periodic re-calibration will be needed.  This can be performed either by the user or by sending the meter to a specialist calibration service or to the manufacturer.  Calibration systems are not simple, and probably only users with multiple instruments will be able to justify maintaining their own calibration systems.

 

Magnetic fields are calibrated by placing them in a known field produced by a coil system of known geometry.  It is common to refer to these coils as a Helmholtz arrangement.  In fact, a Helmholtz pair, two equal coaxial circular coils separated axially by one radius, is a legacy of the time when fields had to be calculated analytically; its main advantage is the ability to express analytically the field and its first few derivatives at the centre.  There are alternative systems which are more efficient at producing a uniform field (for a review, see Kirschvink 1992).  The simplest system is a single square coil; the best results are achieved from three or four coaxial coils, of carefully chosen numbers of turns and axial separation.  The size of the coil is determined by the requirement that the field at the centre must be reasonably uniform over the area of the search coil in the instrument being calibrated.  IEC requires less than 1% (or 1.5% for larger probes under certain circumstances) departure of the field anywhere over the area of the search coil from the field at the centre.  For a 0.1 m diameter search coil, this could be achieved by a single square coil of side 1 m, or a Merrit four-coil system of side 0.4 m.

 

IEC requires the field in the calibration coil to be known to ±3%.  In most practical situations, the geometry of the coil system is known to rather better than this, so the limit on the accuracy of the calibration system is the accuracy of the measurement of the current flowing in the coils.

 

Electric fields are calibrated by placing them between two parallel plates across which a known voltage is applied.  The uniformity of the field is improved by grading rings round the edges of the system, but to avoid perturbation, the system of plates should still be placed a minimum distance away from objects that might perturb the field.  The limit on the size of the plates is then proximity effects between the meter and the plates.  IEC specifies that a meter of maximum dimension 0.23 m requires plates of 1.5 m square separated by 0.75 m.

 

Calibration at high electric fields may involve applying voltages to the calibration plates sufficiently high such that corona becomes an issue and corona guard rings become necessary.

 

For both electric and magnetic fields, when a calibration is performed, there are three alternatives.  One is simple to accept the instrument as passing the calibration provided it is within a certain margin of the correct field.  A second is to adjust the instrument so it reads the correct value, and a third is to record a correction factor which should be applied to any readings taken to give the correct reading.  IEC allows all three, and the choice is determined by the ease of adjusting the meter and the accuracy required.  The option of recording and subsequently applying a correction factor gives the highest accuracy, but is best avoided in many practical measurement situations where it increases the potential for mistakes.

 

All calibrations should be performed as part of a quality-controlled calibration system with traceable records.  Exactly how this is done and what is required depends on the quality regime in operation.

 

Calibrations should be performed at set intervals.  It is common to make this yearly, but if documented experience shows that the meter retains its calibration acceptably for longer than this, as will often be the case with commercial instruments, then a longer calibration interval is acceptable.

6              Sample measurement procedures

This chapter has emphasised that there are no universal rights and wrongs in EMF measurements; the choices made depend on the purpose of the measurements.  The following are therefore suggestions for the particular scenarios identified at the start of this chapter and should not be regarded as definitive.

 

·               Simple characterisation of the field in a building

The objective is to characterise the field in a home, a work location, or similar, by a single number.  Depending on resources and time available, this can be done by a single measurement at a single location; a single measurement at each of a series of locations; a sequence of measurements over time at a single location; or a sequence of measurements over time at multiple locations.

 

Probable measurement procedure: use a battery-powered instrument with logging facilities and three ferrous-cored coils.  Leave it in a standard location for 24 hours or longer.  The location should be clearly specified in the study protocol, eg “1 m above floor level at the middle of the side of the child’s bed”, and should be distant (at least 1 m) from any items of electrical equipment.  This usually means the centre of a room, or as near it as is compatible with the occupants’ use of the home.  Standardisation may be improved by providing a stand to hold the meter. The logging interval will be determined by the need for the logging capacity to last the required time.  An instrument with 1 nT resolution will be preferable in homes but 10 nT resolution may be acceptable.

 

Similar measurements for electric fields are possible but will be more problematical because of perturbations by people walking near to the meter.

 

·               Identification of sources

The objective is to make measurements of the field at a location, specifically how it varies over space or time, to enable the source of the field to be identified.

 

Probable measurement procedure: use a hand-held, battery-powered instrument with a clear read out (analogue will be easier but digital is acceptable as well).  Walk round the location observing the reading so as to identify areas of high field, and track these to their source.  It may be helpful to have a single-axis readout as the direction of the field helps determine the direction of the current producing it.

 

·               Comprehensive characterisation of the field in a building

The objective is to collect more data on the field in a building than just a single number, so as to permit extraction of desired information at a later date.

 

Probable measurement procedure: use several instruments to measure further characteristics of the field.  For example: to characterise the field at more than one location over a house, one approach is to leave several instruments logging for 24 hours at locations spread over the house.  Another approach is to measure a profile of field (using a single instrument but making the measurements as close together as possible) from one side of the house to the other.

 

·               Characterisation of sources

The objective is to perform measurements that relate to a particular source of field (eg a power line, an item of equipment) rather than to the field in a particular place, so as to characterise that source.

 

Probable measurement procedure to characterise a power line: use a three-coil magnetic field meter.  The size of the coils is less important.  Choose one or a few standardised locations, eg directly under the centreline of the power line at mid-span, and possible some standard distances perpendicular to the line, eg 10 m, 25 m, 50 m.  Perform measurements at 1 m above ground.  For electric fields, use a single-axis meter, aligned to measure the vertical field, held on a horizontal insulating pole at 1 m above ground, or a clean, dry vertical insulating stand.  If extra detail is desired, measure the field at more than one height: ground level, 1 m and 2 m above ground.  Measurements at one point in time obviously give the field only at that time.  If the load on the line is available, the measured field at one load can be scaled to other loads; or the measurements can be logging measurements over an extended period of time.

 

·               Personal exposure

The objective is to measure the exposure of a person over a period of time during which they are exposed to fields from various sources.

 

Probable measurement procedure: use a small three-axis meter.  The priority is to prolong battery life and to reduce the bulk, so factors such as the flatness of the frequency response may be sacrificed.  Get the subjects to wear it at a standard position on their body, possibly in a pouch on a belt, or for children, in a child-friendly back-pack or similar.  Preferably arrange for any display to be blanked to reduce the temptation for the subject to experiment with approaching high-field sources.  Give clear instructions on what to do at night time, ie where to place the meter when it is taken off.

 

There are some very specific issues related to personal exposure measurement (dosimetry) for the assessment of occupational exposures to magnetic fields.   They are discussed here.


References

Diplacido J, Shih C H and ware B J 1978 Analysis of the proximity effects in electric field measurements IEEE Transactions on Power Apparatus and Systems PAS-97 2167-2177

 

IEC 1995 Electromagnetic compatibility (EMC) IEC 61000

 

IEC 1998 Measurement of low-frequency magnetic and electric fields with regard to exposure of human beings – special requirements for instruments and guidance for measurements IEC 61786

 

Kirschvink J L 1992 Uniform magnetic fields and double-wrapped coil systems: improved techniques for the design of bioelectromagnetic experiments Bioelectromagnetics 13 401-411